Controller states

StateDescription
IdleNo current activity or lockout
NoFillNo current activity, new fills locked out
VentingVenting dewar 2 in preparation for a fill
FillingFilling dewar 2 from dewar 1
AlarmA datum has caused Venting or Filling to abort

Controller state diagram

Alarm limits

LimitValue
Venting Timeout15 minutes
Filling Timeout40 minutes
Dewar 1 empty weight350 pounds
Dewar 2 full weight600 pounds
Dewar 2 filling pressure10 psi
NoFill start time1400Z
NoFill end time1800Z